Universal bimetallic thermometer

The universal bimetallic thermometer is an industr
  • Tel:18019899982
  • Fax:0550-7989666
  • Mail:tkjd@tiankang.com
Description

Summary:


The universal bimetallic thermometer is an industrial bimetallic thermometer. The main component is a multilayer metal sheet laminated by two or more metal sheets. In order to improve the sensitivity of temperature measurement, metal sheets are usually made into spiral coil shapes. When the temperature of the multilayer metal sheet changes, the expansion or shrinkage of the metal layers is not equal, so that the spiral coil is rolled up or loosened. Because one end of the spiral coil is fixed and the other end is connected to a pointer which can rotate freely, the two metals change differently when the temperature changes, so they will bend. One end is fixed and the other end changes with temperature, and the displacement is close to the temperature. The self recording system is made up of self recording clock and self recording pen. The self recording pen is connected with the magnifying lever and is controlled by the sensing element. When the bimetal feels the temperature change, the pointer can indicate the temperature on a circular indexing scale. The temperature range of this instrument is 200~650 degrees, and it is allowed to be around 1% of the scale two.


Characteristics:


1. The scene shows temperature, intuitive and convenient, reliable and long service life.

2, various structural forms can meet different requirements.

Working principle:

The bimetallic thermometer is based on a bimetallic strip circled and bent. When one end is heated and expanded, the pointer is rotated, and the working instrument displays the temperature value.

Technical parameters:


Implementing standard JB/T8803-1998 GB3836-83

Scale nominal diameter: 60100150

Accuracy: (1), 1.5

Thermal response time: less than 40S

Protection level: IP55

Angle adjustment error: angle adjustment error should not exceed 1% of its range.

Return error: bimetallic thermometer return error should not be greater than the absolute error limit.

Repeatability: bimetal thermometer repeatability limit should not be greater than the basic error limit of the absolute value of the 1/2 main technical parameters.

Product implementation standard: JB/T8803-1998? GB3836-83

Scale nominal diameter: 60100150

Accuracy level: (1), 1.5

Thermal response time: less than 40S

Protection level: IP55

Angle adjustment error: angle adjustment error should not exceed 1% of its range.

Return difference: thermometer return error should not be greater than the absolute error limit.

Repeatability: thermometer repeatability limit should not be greater than the basic error limit 1/2 of absolute value.



Browse: